292x Filetype PDF File size 0.38 MB Source: www.hse.gov.uk
Health and Safety
Executive
Volatile organic compounds
in air
Laboratory method using sorbent tubes, solvent desorption
or thermal desorption and gas chromatography
Scope
1 Three methods are described for the determination of time-weighted average
4 concentrations of volatile organic compounds (VOCs) using air sampling onto
sorbent tubes followed by gas chromatographic analysis. Method 1 uses pumped
sampling with sample introduction by thermal desorption, Method 2 uses diffusive
sampling also with sample introduction by thermal desorption and Method 3 uses
0 pumped sampling and solvent desorption. All three methods are suitable for
personal and static sampling.
2 These methods are applicable to a wide range of VOCs. For Methods 1 and
2, two sorbents are recommended for general use for hydrocarbons, halogenated
hydrocarbons, esters, glycol ethers, ketones and alcohols. For Method 3 the most
common sorbent used is activated coconut shell charcoal although others are
available for specific applications.
Summary
3 A general guide to which of the three sorbent tube monitoring methods is
appropriate for volatile organic compounds dependent upon required sampling time
and estimated analyte concentration is given in Table 1.
4 The upper limit of the useful range of the three methods is determined by the
sorptive capacity of the sorbent used and by the linear dynamic range of the gas
chromatograph column and detector. The lower limit of the useful range depends
on the noise level of the detector and on blank levels on the sorbent tubes or
desorption solvent.
5 Method 1 (pumped sampling; thermal desorption) is suitable for sampling
-3
times of the order of a few minutes (0.1–100 mg.m ) up to 2 hours (low flow rate,
-3
approximate range 0.1–10 mg.m . A measured volume of air is drawn through a
sorbent tube containing appropriate sorbent, specifically selected for the compound
or mixture to be sampled. The collected vapour is then thermally desorbed in an
MDHS1 inert carrier gas into a gas chromatograph fitted with a suitable capillary column
and detector.
6 Method 2 (diffusive sampling; thermal desorption) is suitable for samples of
greater than 15 minutes up to 8 hours over a concentration range of approximately
-3
1–1000 mg.m . The sorbent tube, fitted with a diffusion cap, is exposed for a
measured period of time, and then thermally desorbed in the same manner as
Methods for the Method 1.
Determination of 7 Method 3 (pumped sampling; solvent desorption) is suitable for samples of a
Hazardous Substances few minutes up to 8 hours over a concentration range of approximately 1–1000
-3
Health and Safety mg.m . A measured volume of air is drawn through a sorbent tube, which is then
Laboratory desorbed with a solvent, typically carbon disulphide for simple hydrocarbons.
Health and Safety
Executive
8 In all three methods, the desorbed samples are quantified by gas
chromatography (GC) using either a flame ionisation (FID), mass selective (MS), or
other suitable detector.
9 The use of alternative methods not included in the MDHS series is acceptable
provided they can demonstrate the accuracy and reliability appropriate to the
application.
Recommended sampling
10 Method 1 (pumped sampling; thermal desorption): recommended sampling
-1
rate: 50–100 ml.min ; typical sampled volume: 1 to 10 litres; maximum sampled
volume: see Tables 2a, 2b and 3 for details corresponding to Tenax TA, Carbopack
X and Chromosorb 106 sorbents respectively.
11 Method 2 (diffusive sampling; thermal desorption); minimum sampling time:
15 – 30 minutes (up to 8 hours); see Table 4 for details of diffusive uptake rates for
a range of analyte/sorbent combinations.
12 Method 3 (pumped sampling; solvent desorption); recommended sampling
rate: 50 – 200 ml.min-1; typical sample volume: 10 litres.
Prerequisites
13 Users of Methods 1 and 3 where pumped sampling is required will need to be
1
familiar with the content of MDHS14.
Safety
14 Users of this method should be familiar with standard laboratory practice and
be able to carry out a suitable risk assessment. It is the user’s responsibility to
establish appropriate health and safety practices and to ensure compliance with
regulatory requirements.
Equipment
Sorbent tubes
Methods 1 and 2
15 Pre-packed stainless steel sorbent tubes. Typically, the tubes should be
packed with 200 mg to 300 mg of a suitable sorbent (or sorbents) of particle size
0.18 mm to 0.5 mm (35 to 80 mesh). The sorbent is usually retained between two
steel gauzes. Each tube should have metal storage end caps fitted with PTFE
seals. The inlet end of the tube is usually marked by a scored ring about 10 mm
from the end. Ideally the tube should have a unique engraved/etched identification
number or bar code. Avoid the use of adhesive labels on the body of the tube that
can interfere with the tube transfer mechanism in the thermal desorption
equipment.
16 Diffusion end caps (Method 2 only). Standard end caps contain a metal gauze
only; special caps with an additional silicone membrane may be used to reduce
water uptake when using hydrophilic sorbents.
Volatile organic compounds in air Page 2 of 30
Health and Safety
Executive
17 Select a sorbent tube appropriate for the compound or mixture and the
required sampling duration. Guidance on selection of suitable sorbent tubes is
2
given in Tables 2a, 2b, 3 and in EN ISO 16017-1 for active sampling and in table 4
3
and EN ISO 16017-2 for passive sampling. For active sampling, care should be
taken to avoid breakthrough of the analyte and the recommended safe sampling
volume (SSV) should be implemented. See Appendix 1 for further information on
breakthrough volume and SSVs.
18 The industry standard stainless steel sorbent tube dimensions are typically
either 3.5” (89 mm) long × ¼” (6.4 mm) OD × 5 mm ID or 7” (178 mm) long × 6
mm OD × 5 mm ID (not used passively, glass tubes with similar dimensions are
also available ). The air gap dimensions (distance between stainless steel screen
retaining the sorbent bed and the inlet end of the tube) which is fundamental for
accurate diffusive sampling should be in the range 14.0 mm to 14.6 mm. The
diffusive uptake rates given in Table 4 are based on the 3.5” stainless steel tube.
Additional information on the tube construction, dimensions, range of sorbent types
2 3
and applicability are given in EN ISO 16017-1 or EN ISO 16017-2.
19 The sorbent tubes should be conditioned according to the manufacturer’s
instructions before use, either on the instrument conditioning cycle or alternatively
under a flow of dry inert gas such as nitrogen at typically 100 ml min-1 for 10
minutes with gas flow in the opposite direction to that used for sampling. The tube
should be heated to a temperature usually at least 25 °C below the manufacturer’s
recommended maximum temperature for the sorbent but above the analysis
desorption temperature and recapped immediately after cooling. Guidance on
2
typical conditioning parameters for common sorbents is given in EN ISO 16017-1
3
or EN ISO 16017-2.
20 For new tubes, after conditioning, a representative sample, (typically 10%)
should be analysed to ensure that the thermal desorption blanks are acceptable.
The sorbent tube blank level should be very low (eg <2 ng) but with use of high
sensitivity detection methods very low sample amounts can be analysed. Rather
than stating absolute values, it may be considered acceptable if interfering artefact
peak areas are less than 5% of the typical area of the analyte(s). If levels of
artefacts exceed these levels, reconditioning should be considered.
21 The desorption efficiency of the analyte from the sorbent tube will be
quantitative (>95%) provided that appropriate desorption parameters are adopted.
Typical desorption temperatures are given in Tables 2a, 2b and 3 for a range of
organic compounds on Tenax TA, Carbopack X and Chromosorb 106 sorbents
2,3
respectively. Information on additional sorbents can be found elsewhere. The
efficiency of desorption can be checked by analysing a clean empty glass tube
(to identify that the analyte has been fully desorbed from the trap) after the sorbent
tube, followed by reanalysis of the first tube to confirm quantitative desorption has
been achieved by the first analysis. This should be repeated for a range of loadings
to cover the calibration range of the method. Alternative procedures to determine
2,3
desorption efficiency are described elsewhere.
Method 3
22 A wide variety of pre-packed sorbent tubes for solvent desorption is
commercially available, with coconut shell charcoal amongst the most widely used.
Guidance on selection of suitable sorbent tubes is given in Table 5. Typical tubes
are made of glass 70 mm to 110 mm long with an OD of 6 mm to 8 mm. The
tubes are generally supplied with both ends flame sealed and usually containing
Volatile organic compounds in air Page 3 of 30
Health and Safety
Executive
two sorbent beds, separated and held in place by glass wool or foam plugs. In the
case of coconut shell charcoal, the front (sample) bed usually contains 100 mg of
sorbent and the back-up bed 50 mg. When sampling, glass tubes should be held
in a protective holder when sampling in order to prevent breakage.
23 There may be some batch-to-batch variation in sorptive capacity of coconut
shell charcoal. Tubes from a single batch should ideally be used in each
application. Sorbents other than charcoal are available and may be used for certain
applications.4
24 The desorption efficiency (D) of sorbents may vary for type, batch and mass
loading. It is therefore necessary to determine D for each sorbent and analyte over
4
the sample concentration range. See Appendix 2 or BS ISO 16200-1 for
procedures describing determination of D.
25 The sorbent tube sampling efficiency will be 100%, provided that the sampling
capacity of the sorbent has not been exceeded. If this capacity is exceeded,
breakthrough of vapour from the front section to the back-up section will occur.
The breakthrough volume varies with ambient air temperature, relative humidity,
concentration of sampled vapour and of other contaminants, and with the sampling
flow rate. An increase in any of these parameters causes a reduction in the
breakthrough volume (See Appendix 1 for further information on breakthrough
volume). The back-up section may be used as a check on breakthrough under
practical conditions.
26 Published methods and validation data giving detailed information on
breakthrough volumes, sampling flow rates, sampling times and analysis for
5
specific analytes are available from the US regulatory authorities, ie OSHA and
6
NIOSH.
Other equipment
7
27 Personal sampling pumps that meet the requirements of BS EN ISO 13137.
28 A portable flow meter calibrated against a primary standard, with a
measurement uncertainty typically less than ±2%.
29 Flexible plastic tubing for making a leak-proof connection from the sorbent
tube to the pump; belts or harnesses to facilitate attachment of sampling apparatus
to sample subjects.
Laboratory reagents and apparatus
30 During analysis use only reagents of a recognised analytical quality.
31 Volatile organic compounds: for preparation of calibration standards.
32 Methanol: diluent for the preparation of sorbent tube standards (Methods 1
and 2; alternative solvents can be used when appropriate).
33 Carbon disulphide: for solvent desorption of sorbent tubes (Method 3)
34 A calibrated 5 µl syringe capable of dispensing accurate volumes between 1
µl and 5 µl: for preparation of spiked sorbent tube standards (Methods 1 and 2).
35 A range of volumetric flasks with caps or suba-seal closures, Class A
8
complying with the requirements of BS EN ISO 1042.
Volatile organic compounds in air Page 4 of 30
no reviews yet
Please Login to review.